7.7. Cleaning the Probe
The probe tube and gas sample hose will become dirty under normal use.
NOTE: T
he water trap’s filter element should prevent soot from
reaching the analyzer’s internal components. If the probe is not kept
clean, it could become clogged and restrict the flow of gas into the
analyzer, resulting in incorrec
t combustion test readings and
calculations.
NOTE:
An analyzer that is used to test natural gas furnaces
normally requires less frequent cleaning than an analyzer used for
testing coal or oil fired furnaces.
7.7.1. Equipment Required
Alcohol
Aerosol Can of Automotive Carburetor Cleaner
Clean Rag
Source of Compressed Air (optional)
7.7.2. Procedure
1. Remove gas sample hose from top of water trap.
CAUTION:
Carburetor cleaner attacks plastic
components! Take precautions not to spray cleaner onto
the probe handle or analyzer.
2. Insert the plastic spray tube of the carburetor cleaner into the gas
sample hose, and then liberally spray carburetor cleaner through the
hose and out the probe tube.
3. After spraying, remove all the residual cleaner by repeatedly flushing
the gas hose and probe tube with alcohol.
4. Wipe off the surfaces of the probe and tubing with a clean rag.
5. Allow the parts to dry completely. If available, blow compressed air
through the probe to expedite the drying process.
6. Reconnect gas sample hose to top of water trap.