Bacharach PCA3 Spezifikationen Seite 43

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 104
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 42
PCA
3 Manual
0024-9473 Rev 4 43
4. Use the ▲▼ buttons to highlight RESET FORMAT, then press ENT to
display the Reset Format screen.
5. Use the ▲▼ buttons to select YES, then press ENT to reset the display
and return to the Run/Hold Format screen.
Figure 4-25. Resetting Display Formats Back to Factory Defaults
4.22. NO
x
Calculation
NO
x
is a combination of NO and NO
2
. The actual NO
x
value on the PCA3 is the
sum of both NO and NO
2
readings (NO
x
=NO+NO
2
) from the sensors in positions
4 and 3, respectively. Analyzers having just the NO sensor (in sensor position
#4) can display a
calculated
NO
x
approximation using just the NO sensor reading
and a multiplier of 1.05 (NOx
NO*1.05). This calculation is enabled by setting
the NOx Calculation option in the Setup Menu to ON.
Figure 4-26. NOx Calculation Screens
Figure 4-27. Sensor Configurations for Actual and Calculated NOx Values
Seitenansicht 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 103 104

Kommentare zu diesen Handbüchern

Keine Kommentare