PCA
3 Manual
0024-9473 Rev 4 33
4. Use the ▲▼ buttons to highlight the desired measurement, and then
press ENT to display the REF TO O
2
screen for that measurement. In
the example shown, CO has been selected.
5. First use the
◄►
buttons to move the cursor across the screen until it is
over the digit to be changed, and then press the ▲▼ buttons until the
desired value is displayed.
6. Press ENT to save the displayed value and re-display the O
2
REFERENCE screen, or press ESC to abort this procedure and retain
the old O
2
reference value.
7. If the O
2
reference value for more than one gas is being set, repeat
Steps 4, 5, and 6 for each measurement.
Figure 4-12. O
2
Reference Setup
4.13. Print Pressure Selection
Select whether to print or not print the pressure measurement on the combustion
test printout as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight PRINT PRESSURE, and then press
ENT to display the PRINT PRESSURE screen.
4. Use the ▲▼ buttons to highlight either No (do not print pressure) or Yes
(print pressure). In the example shown, Yes has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.