72 0024-9473 Rev 4
Alternatively:
Plug the probe’s thermocouple into the T-AIR connector
located at the bottom of the analyzer. DO NOT attach the probe’s gas
hose to the analyzer’s GAS port; otherwise water will be drawn into the
analyzer!
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 66).
3. Use the ▲▼ buttons to highlight T-Air, and then press ENT to display
the CALIBRATE TA-ZERO screen.
NOTE: “Measured” is the current temp
erature reading,
while “Applied” is a known temperature that will be applied
for calibration purposes.
4. Set thermocouple simulator to 0 °C (32 °F), and then use the ▲▼ and
◄► buttons to enter an applied value that exactly equals the setting of
the simulator.
Alternatively:
Submerge probe tip into an ice-water bath with a
thermometer, wait several minutes, and then use the ▲▼ and ◄►
buttons to enter an applied value that exactly equals the thermometer
reading.
NOTE: The calibration range is from 0 to 5 °C (32 to
41 °F
). An attempt to calibrate outside this range will
cause the message “Applied Value High” (or Low) to
appear at the bottom of the screen.
5. Wait until the measured reading stabilizes, and then press ENT to
calibrate the TA-Zero Measured value to that of the applied value; after
which the message “Good Calibration” should briefly appear followed by
the CALIBRATE TA-SPAN screen.
Figure 6-5. T-Air Zero Procedure