Field
Column Name
Data Name or Value
18 TempUnit Temperature Units
19 NO Nitric Oxide Level (see field 32)
20 ppm NO(O
) Nitric Oxide Level referenced to a % of O
(see field 28)
21 NO
2
Nitrogen Dioxide Level (see field 32)
22 ppm NO
2
(O
2
) Nitrogen Dioxide Level referenced to a % of O
2
(see field 28)
23 NO
x
Nitrogen Oxides (NO+NO
2
) (see field 32)
24 ppm NO
x
(O
2
) Nitrogen Oxides Level as referenced to a % of O
2
(see field 28)
25 SO
Sulfur Dioxide (see field 32)
26 ppm SO
2
(O
2
) Sulfur Dioxide Level referenced to a % of O
2
(see field 28)
27 CO_O
2
Ref %O
2
Reference for CO
(1)
28 NO
x
_O
2
Ref %O
2
Reference for NOx
(1)
29 SO
2
_O
2
Ref %O
2
Reference for SO
2
(1)
30 Pressure Pressure (Draft) Value
31 Pressure Units Pressure Units of Measure
(2)
32 Pollution Units Pollution Units of Measure
(3)
33 NO Temp Nitric Oxide Sensor Temperature
34 PS Voltage Power Supply Voltage
(1) %O
2
reference as selected (refer to O2 Reference Setup on page 32)
(2) Pressure units as selected (refer to Pressure Units Selection on page 27)
(3) Pollution units as selected (refer to Pollution Units Selection on page 28)
5.16. Printing Test Data
Combustion or pressure data that is currently being displayed can be sent to a
printer using IrDA protocol as described below.
Data that is stored in memory can also be printed by first displaying the stored
test data as described in Recalling Combustion Test Data section (page 58) and
Recalling Logged Test Data section (page 59).
In addition to printing combustion and pressure data, the contents of any screen
that shows the label “PRINT” above the
F1
button can be printed. For example,
the information shown in the DIAGNOSTIC screen can be printed.
1. Turn ON printer. Refer to the printer’s instruction manual for detailed
operating information. If not already done, set up the printer for:
•
8
bit
•
No parity
•
9600
baud
•
IrDA is set to IrDA-SIR
•
DTR handshaking.
2. Align the printer with the top of the analyzer as shown in Figure 5-23.
3. Press the PRINT (F1) button to begin printing.