Bacharach Fyrite INSIGHT PLUS Bedienerhandbuch Seite 92

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 116
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 91
Calibration and Maintenance Fyrite
INSIGHT
Plus Manual
86 0024-9487 Rev 2
5.5. Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
5.5.1. Materials Required
Bellows
Manometer
Range: ± 8 in. of water column (± 20 mB)
Accuracy: ± 0.01 in. of water column (± 0.025 mB)
5.5.2. Procedure
NOTE: The unit-of-measure for pressure is selected from the Pressure
Units parameter in the Setup Menu. In the following procedure “inwc
(inches water column) is selected, but note that any unit-of-measure can
be used for calibration purposes. Below are unit conversions for
reference.
249 Pascals/inwc
2.49 mB/inwc
2.49 hPa/inwc
25.4 mm H
2
O/inwc
Seitenansicht 91
1 2 ... 87 88 89 90 91 92 93 94 95 96 97 ... 115 116

Kommentare zu diesen Handbüchern

Keine Kommentare