Configuration Fyrite
INSIGHT
Plus Manual
52 0024-9487 Rev 2
Setup Menu Function
The COMBUSTION EQUATIONS menu allows the user
to select either Siegert combustion equations or North
American combustion equations.
Use UP () and DOWN () arrow buttons to highlight
the desired option. Press ENTER to select.
IMPORTANT: Changing this setting resets
several configuration parameters to their
default values. Below is a list of affected
parameters and those unaffected.
Reset to
Default Values
Unchanged
Temperature units Manual/Auto zero
Pressure units Calibration data
O
2
(Oxygen) reference User name
Print pressure Test ID
Zoom O
2
sensor type
Button sound Clock
RUN/HOLD format
Fuel
Memory erased
The CO Protect feature protects the CO sensor from
the negative effects of being overloaded with CO gas.
Such negative effects include:
a longer sensor recovery time
a shortened sensor life.
The CO sensor is an electro-chemical sensor whose