Monoxor ® IIICarbon Monoxide AnalyzerInstruction 19-9326Operation & Maintenance Rev. 2 – October 2003Product Leadership • Training • Service • Rel
Instruction 19-93266Monoxor III3.2 Probe InstallationInstall the probe if the analyzer is going to be used for ue-gas CO mea-surements, or if gas me
Instruction 19-93267Monoxor III3.3 Front Panel PushbuttonsNote that a push button may perform several functions, depending on what screen is being di
Instruction 19-93268Monoxor III3.4 Setup ModeThe analyzer is preset at the factory for the parameters shown below, but can be changed as described i
Instruction 19-93269Monoxor III1. Enter the Setup Mode per Section 3.4.1. If necessary, repeatedly press ENTER until the Temperature Units Setup Scre
Instruction 19-932610Monoxor III3.4.4 Time SetupThere are two Time Setup Screens, one for hours and the other for min-utes. Two bars appear above th
Instruction 19-932611Monoxor III3.4.6 Year SetupThe Year Setup Screen is labeled “yEAr”.1. Enter the Setup Mode per Section 3.4.1. If necessary, r
Instruction 19-932612Monoxor IIINotes
Instruction 19-932613Monoxor III4.0 OPERATIONTo operate the Monoxor III, you simply . . .• Turn the analyzer ON• Wait for the unit to warm up• Take a
Instruction 19-932614Monoxor III4.2 Ending a TestWARNING! Burn Hazard. Do not touch the probe after removing it from a ue. Allow the probe to cool
Instruction 19-932615Monoxor III4.5 CO Sensor Error ScreenWhen the CO channel is set up for Auto Zero (refer to Section 3.4.3), a CO sensor error wil
WARRANTYBacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufature and will on
Instruction 19-932616Monoxor III4.8 Operating Tips• When an analyzer is brought in from a cold vehicle, let it warm up slowly to minimize condensatio
Instruction 19-932617Monoxor III4.10 Printing Test ResultsTurn ON the printer. Refer to the printer’s instruction manual for detailed operation and m
Instruction 19-932618Monoxor IIINotes
Instruction 19-932619Monoxor III5.0 CALIBRATION & MAINTENANCEImportant: Fresh batteries should be installed, and the unit allowed to stabilize at
Instruction 19-932620Monoxor III5.2 Ambient Temperature CalibrationMaterial Required: Calibrated ThermometerProcedure:1. Enter the Calibrate Mode
Instruction 19-932621Monoxor IIIFigure 5-1. Calibration Kit Hookup
Instruction 19-932622Monoxor III5.4 Water Trap / Filter MaintenanceThe Water Trap / Filter Assembly removes water condensate from the gas sample, and
Instruction 19-932623Monoxor III5.5 CO Sensor ReplacementMaterial Required: • CO Sensor, P/N 24-7265 • CO Sensor Gasket, P/N 24-1112 • #1 Phillips
Instruction 19-932624Monoxor III5.6 Pump Assembly ReplacementMaterial Required: • Pump, P/N 24-3020 • #1 Phillips ScrewdriverProcedure:1. Disassem
Instruction 19-932625Monoxor IIIFigure 5-3. CO Sensor Located on Backside of Printed Circuit Board
Monoxor IIIInstruction 19-9326iContents1.0 INTRODUCTION ...12.0 TECHN
Instruction 19-932626Monoxor IIIFigure 5-4. Back Case Component Locations
Instruction 19-932627Monoxor III6.0 PARTS & SERVICE6.1 Replacement PartsDescription Part No.Carbon Monoxide Sensor ...
Instruction 19-932628Monoxor III6.3 Service CentersReplacement parts and service can be obtained by contacting one of the fol-lowing Bacharach Servic
Instruction 19-932629Monoxor IIIFor additional information on CO and its dangers, ask your sales represen-tative about Bacharach’s Institute of Techni
Monoxor IIIInstruction 19-932630Notes
Monoxor IIIInstruction 19-932631Notes
World Headquarters621 Hunt Valley Circle, New Kensington, PA 15068Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666Website: www.bacharach
Monoxor IIIInstruction 19-9326iiNotes
Instruction 19-93261Monoxor III1.0 INTRODUCTIONThe Monoxor III is a portable analyzer designed to detect and display concentrations of Carbon Monoxid
Instruction 19-93262Monoxor IIINotes
Instruction 19-93263Monoxor III2.0 TECHNICAL CHARACTERISTICSCO Display Range ... 0–2000 ppmAccuracy ...
Instruction 19-93264Monoxor IIINotes
Instruction 19-93265Monoxor III3.0 PREPARING THE ANALYZER FOR OPERATIONTo prepare a new analyzer for operation, you must install four ‘AA’ Alkaline