The Informant™ 2
0019-9211 Rev 9 19
4. TROUBLESHOOTING
The following table lists the causes and remedies for the most common
problems that may occur with the detector.
If the information in Table 1 does not solve the problem, or for help with
any problem that is not listed, please contact one of the Bacharach
Sales/Service Centers listed in Section Parts & Service.
Table 1. Troubleshooting Guide
Symptom
Instrument does not
turn on.
backward. Install four fresh ‘AA’
alkaline batteries per Section Battery
Installation.
b. Possible faulty refrigerant sensor.
-Remove refrigerant sensor and turn on
detector. If detector now turns on, and
after warm-up shows a bad sensor (see
sensor failure indication below), replace
sensor per Section Sensor
Replacement.
c. Loose printed circuit board connector.
Disassemble detector and ensure that
all connectors are securely attached to
The middle gas-level
LED on both displays
glows steady.
Section Sensor Replacement.
All gas-level LEDs
momentarily light full
scale and then turn
off, leaving both the
audible tick rate and
probe tip blink rate at
-elevated levels.
refrigerant. Move the detector to an area
of fresh air for about 1 minute, or until all
indicators return to their idle state.